Your browser does not support JavaScript!

SAT 450

HOW DOES IT WORK?

 

SAT is a fully automatic, random access analyzer designed to support the modern laboratory in achieving improved cost efficiency, reduced rerun and manual handling.

It provides quality results while following rigid quality system procedures and in strict compliance with regulatory agency requirements. 

 

 


SPECIFICATION

SYSTEM DESCRIPTIONFully Automatic, random access, continuous loading, benchtop open analyzer for Clinical Chemistry and Immunoturbidimetric assays.
ASSAY TYPESEnd point, Kinetic, Differential, Fixed time, Bichromatic and Immunoturbidimetric. Up to 999 tests, all user programmable
REAGENTS ON BOARD

Up to 72 reagents can be located into the refrigerated compartment and distribuited on:

- 4 removable racks

- 50/20/5 ml containers can be loaded

- 8 extra positions for 5 ml containers

SAMPLE LOADING

- 4 universal, separate racks for continuos loading of samples, calibrators and controls

- Each rack contains up to 15 cups and/or tubes 10-16 mm diam., 40-100 mm height

- 8 extra position for STAT samples, calibrators and controls

- STAT sample execution is allowed at any time

THROUGHPUTUp to 420 tests/hour including I.S.E. (280 tests/hour photometric).
READING SYSTEM

- Direct reading, dual channel photometer

- 200 μl minimum reading volume

- 9 narrow band, automatically selected interferential filters 340, 380, 405, 492, 510, 546, 577, 620, 690

- Halogen lamp 6V/10W

- Linear from 0.0001 to 4.200 Abs.

- High resolution (0.0001 Abs.)

SAMPLING PROCESS

One mechanical arm performs all sampling operations with:

- Liquid level sensing (capacitive)

- Reagent pre-warming at 37 °C

- Automatic probe washing

- Sample volume range: 2.0 μl ÷ 99 μl (0.25 μl incr.)

- Reagent 1 volume range: 2.0 μl ÷ 348 μl (1 μl incr.)

- Reagent 2 volume range: 2.0 μl ÷ 330 μl (1 μl incr.)

- Reagent 3 volume range: 2.0 μl ÷ 330 μl (1 μl incr.)

- Sample pre-dilution

- Sample post-dilution

- Sample post-concentration

REACTION PLATE

Holds 80 cuvettes divide into 4 racks with 20 cuvettes each and performs the following operations:

- Automatic cuvette washing

- Continuous monitoring of cuvette quality

- Incubation temperature control at 37 °C ± 0.3

OPTIONS

- Direct potentiometry I.S.E. module for Na+, K+, CI- (140 tests/hour)

- Positive barcode readers for reagents and samples (code 128, codebar, code 2 of 5 interleaved, code 39)

 

Brand: